The Ventbuster® - Software Update

Software Upgrades 


We would like to share helpful information to enhance the Ventbuster® User experience. As our client, you are receiving regular emails with tips and tricks for operating the Ventbuster®. If you have any questions or feedback, please contact us.

To provide the best possible user experience, we have enhanced our current software platform for the Ventbuster®. Please review the following information and install a software update to both the Ventbuster® Tablet and Ventbuster® Comms Unit:

Ventbuster® Tablet: Version v 0.3.16
Version 0.3.16 of the android application is now available for download from the Google Play Store. The update is private for Ventbuster clients only - please contact us, should you need the update link.

This update will improve the reliability of data uploads from the Ventbuster® Tablet. In a few cases, the Tablet could overrun the rate at which messages could be accepted by the cloud service. Users would potentially see duplicate “Test START” and “Test END” notifications.

Ventbuster® Comm Unit: Version v 1.1.0
Version 1.1.0 of the Comms unit firmware is now available for an over-the-air update. It will improve or correct for four (4) functions:

  • In rare cases, measurements could have a minor timestamp variance. Data could be timestamped from a fractionally earlier period in the test, resulting in small gaps in test data where the measurements should have been.

  • After deleting measurements using the Comms Unit menus or Tablet, in the odd case, the device could “re-discover” previously transmitted measurements and flag them as new un-transmitted measurements.

  • A pointer in the memory that referenced the stored address for the next data transmission was always written to the same flash location. Over time, this could lead to flash wear-out. This problem has not been seen in the field, but software programming has been changed as a precaution.

  • With the upgrade, the Ventbuster® Comms unit no longer needs to scan its memory for measurements each time it boots up. This will result in a start-up when there is a large measurement backlog.

Ventbuster® Comm Unit: Version v 1.1.0 Installation
Installing v 1.1.0 will delete any untransmitted measurements stored on the Comms unit. As long as there are un-transmitted measurements, the update will not be installed automatically. To install v 1.1.0 on a device with un-transmitted measurements, choose one of the three options below:

  1. Wait for the Ventbuster® Comms Unit to transmit all measurements directly.

  2. Transfer all measurements to the tablet, then delete them off of the Ventbuster® Comms Unit.

  3. Manually install the update via the user menus, which will delete any un-transmitted measurements.
    * This is only recommended if there is no value to keeping saved any measurements saved on the device.


As always, please refer to the Ventbuster® Field Guide and/or the Ventbuster® Start-up Checklist to set-up, start, and stop any SCVF Test.